Datascope Corp. CS300 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-3023-53 |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during schedule pm by getinge fse the cs300 intra aortic balloon pump (iabp) compressor making loud rattling noise.There was no patient involvement or adverse event reported.The fse replaced noisy compressor.All functional and safety checks to meet factory specifications were performed.
|
| |
|
Additional Manufacturer Narrative
|
|
A getinge field service engineer (fse) replaced noisy compressor.All functional and safety checks to meet factory specifications were performed.The failure analysis and testing dept.Received compressor with a reported unit failure of the compressor making a loud rattling noise.The fat performed a visual inspection and found the part to be in good condition.The fat installed the part into cs300 test fixture and tested the part to factory specifications per the cs300 service manual.Confirmed that the part would make a rattling sound during operation.Since this is a very old compressor, the probable root cause is expected wear and tear.Retaining the part in the failure analysis and testing department.
|
| |
|