Zimmer Surgical SA X SERIES ASEPTIC BATTERY HOUSING; Battery, Replacement, Rechargeable Back to Search Results
Catalog Number 89-8521-470-40
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the battery housing could not be closed due to damaged lid rubber.No consequences or impact to the patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4) g2: foreign - event occurred in iceland.Review of the most recent repair record determined visual inspection revealed damaged lid rubber, lid locking system test confirmed that the device could not close.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES ASEPTIC BATTERY HOUSING
Common Device Name
Battery, Replacement, Rechargeable
Manufacturer (Section D)
Zimmer Surgical SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
Zimmer Surgical SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ   CH-1228
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50943022
Report Number0008031000-2026-00059
Device Sequence Number15827368
Product Code MOQ
UDI-Device Identifier00889024581746
Combination Product (Y/N)N
Initial Reporter CountryIC
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number89-8521-470-40
Device Lot Number5014689
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/15/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-