Model Number 97800
Medical Device Problem Codes Electromagnetic Interference (1194); Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a patient (pt) who was implanted with an implantable neurostimulator (ins) for urinary/bowel dysfunction.It was reported that they noticed today that the communicator was not responding and they didn't have a way to turn off their therapy as they pass through airport security gates.The pt mentioned the communicator would show a blinking yellow light for 60 seconds then would turn off.The pt stated they tried charging the communicator, but it didn't show any indicator lights.The pt added their programmer takes time to charge but was still working and had 67% battery life.The agent reviewed general therapy and device-use information and agent sent repair request to replace the communicator.The agent sent expedited delivery for the replacement.The agent called pt back to gather additional event information, pt answered and provided information.
 
Additional Manufacturer Narrative
Continuation of d10: product id tm90q0 (serial: (b)(6)); product type: 0001-accessory; implant date n/a; explant date n/a medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
InterStim X
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50943077
Report Number3004209178-2026-16577
Device Sequence Number18754939
Product Code EZW
UDI-Device Identifier00763000892067
UDI-Public(01)00763000892067(17)260728(20)01
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2026
Device Model Number97800
Device Catalogue Number97800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured02/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Patient Age61 YR
Patient SexUnknown
-
-