Based on information received from the customer (the executive director), during a transfer using a maxi move lift and clip sling, the right leg clip became detached, resulting in the resident falling from the sling and striking their head.A hematoma was identified during the initial assessment, and the resident was transferred to the hospital for further evaluation.The resident passed away the following day.The reported cause of death was intracranial bleeding.The customer report indicated a possible right leg spasm at the time of the incident.
|
|
The device is distributed outside the us and no gudid information exists.Following the event, the lift and sling (model maa4100m-l, serial number (b)(6)) were inspected by an arjo technician.No malfunctions were identified, and all functions were found to be operating as intended.Based on the information collected, the resident was a full quadriplegic, non-verbal, and had no muscle tone, although occasional spasms had been reported.The investigation is ongoing.Results of the analysis will be provided to the final report.
|