• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (Boxborough) O-Arm O2 Imaging System; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (Boxborough) O-Arm O2 Imaging System; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Medical Device Problem Codes Mechanical Problem (1384); Unintended Electrical Shock (4018)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the door winch cable grade was a c or d and required replacement.(door winch cable grade c/d).Confirmed reported complaint, "door winch cable grade c/d." cable failed visual inspection.Cable was showing signs of wear and fraying.Worn cable.There was no patient involved.
 
Additional Manufacturer Narrative
(h3,h6) : manufacturing representative went to the site to test the system.Fse along with other fse replaced door winch as per asm.Fse found the cable grade to be "d" therefore fse replaced door winch.System functioning as designed.Codes b01, c07, d02 are applicable.(h3, h6): the bi71000429 was returned to the manufacturer for evaluation.The reported issue was confirmed."door winch cable grade c/d." cable failed visual inspection.Cable was showing signs of wear and fraying.Worn cable.Codes: b01, c0201, d02 are applicable.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000429, serial/lot #: w1612011137 rev.8 , ubd: , udi#: ; product id: bi71000273, serial/lot #: -, ubd: , udi#: medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-Arm O2 Imaging System
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough MA 01460
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50943457
Report Number3004785967-2026-00524
Device Sequence Number15840855
Product Code OWB
UDI-Device Identifier00643169639683
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11...."
Patient SexUnknown
-
-