| Model Number |
0998-00-0800-55 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 - event site postal code: 1060
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that during a routine check before use, the cardiosave intra-aortic balloon pump (iabp) had fault #14 (power-up test fails # 14 (prior compressor failure etf)).There was no patient involvement, no harm was reported.A getinge field service engineer (fse) reported that the compressor is planned to be replaced, the error code directly indicates a compressor fault.
|
| |
|