MAUDE Adverse Event Report: Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System Back to Search Results
Model Number 9510-007
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that a no readings/ sensor error/ brief sensor issue occurred.Performance data was reviewed.No readings/ sensor error/ brief sensor issue was not found within the investigation window.The allegation was not confirmed.However, signal loss over an hour was found in connection to the reported allegation.The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dexcom G7 15 Day Continuous Glucose Monitoring System
Common Device Name
Continuous Glucose Monitor
Manufacturer (Section D)
Dexcom, Inc.
6340 sequence dr.
san diego CA 92121
Manufacturer (Section G)
Dexcom, Inc.
232 s dobson rd
mesa AZ 85202
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
+1(858)200-0200
MDR Report Key50943819
Report Number3004753838-2026-263511
Device Sequence Number13136483
Product Code QBJ
UDI-Device Identifier00386270004741
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9510-007
Device Catalogue NumberSTP-FT-013
Device Lot Number1726102002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/15/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/01/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight215 LB
-
-