The customer reported at least 6 picc packs where the secureportiv had been activated and appeared already popped and unable to be used.A sample was received; medline provided the secureportiv samples to the vendor for evaluation, reviewed the dhr and pack build for the kit, and the vendor reviewed production records.The complaint was confirmed.Based on the vendor's evaluation of the sample as well as their manufacturing process for the component, they confirmed that a root cause for the reported issue is most likely due to equipment maintenance; preventative actions have been implemented to address the issue, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|