Covidien EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Catalog Number AB35W07100080
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during post-dilation of a moderately calcified, minimally tortuous distal right common iliac artery lesion with an 8 mm vessel diameter, an evercross 035 pta balloon, burst at 6 atm during the first inflation.The balloon was advanced through a previously deployed everflex 8 mm x 100 mm x 80 mm stent over a.035 non medtronic guidewire and inflated using a non medtronic inflation device with 50/50 contrast.The instructions for use were followed without issue.The balloon was safely removed without detachment, and a new evercross 035 pta balloon, was used.During the first inflation at 6 atm, the second balloon burst.The second balloon was safely removed without detachment, and the physician ended the procedure.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERCROSS 035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
Covidien
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
Covidien
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50944353
Report Number2183870-2026-00479
Device Sequence Number18754839
Product Code LIT
UDI-Device Identifier00763000766788
UDI-Public(01)00763000766788(17)290116(10)C256114(20)10
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K082579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAB35W07100080
Device Lot NumberC256114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured01/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR