Olympus Winter & Ibe GmbH Video telescope "ENDOEYE HD II", 5.4 mm, 0°, autoclavable; rigid video laparoscope
|
Back to Search Results |
|
| Model Number |
WA50050A |
| Medical Device Problem Codes |
Break (1069); Contamination (1120); No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the olympus rigid video laparoscope can see through it during an unspecified procedure.During the initial evaluation of the device, it was observed, fiber breakage, brown stains under light guide cover glass and absolutely no image.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: can see through it, no image and fiber breakage, cause traced to component failure.Brown stains under light guide cover glass, the cause is not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|