The customer observed falsely elevated creatinine results generated on the architect c4000 instrument for a.The following data was provided:
(patient normal ranges: adult male 0.73-1.18 mg/dl, adult female 0.55-1.02 mg/dl)
patient 1
(b)(6) 2026, 22-year-old female
sid (b)(6) initial result 5.87 mg/dl (on (b)(6) repeated 0.73 and 0.69 mg/dl (the patient sample was sent to a sister hospital and repeated on two separate alinity c analyzers).The patient was sent to emergency room at sister hospital, and a sample was drawn that yielded normal results.The patient did not receive any treatment.Patient 2
(b)(6) 2026, 29-year-old female
sid (b)(6) initial result 2.02 mg/dl (on (b)(6), repeated 0.61 and 0.57 mg/dl (patient sample was sent to a sister hospital and repeated on two separate alinity c analyzers)
no impact to patient management was reported.
|