| Model Number |
K10001141 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported constantly leaking water, and the amber light was indicating that there was an internal issue during inspection for use involving an olympus flushing pump.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer has been confirmed.A root cause could not be identified.However, based on the results of the investigation, the most probable cause of the constantly leaking water and the amber light indication of an internal issue was that the pump head did not close tightly on the tubing.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|