Olympus Winter & Ibe GmbH Video telescope "ENDOEYE HD II", 10 mm, 30°, autoclavable; Laparoscope, Gynecologic
|
Back to Search Results |
|
| Model Number |
WA50042A |
| Medical Device Problem Codes |
Crack (1135); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported poor image quality involving an olympus laparoscope, gynecologic.Additionally, during the initial evaluation of the device, a green image and a cracked cover glass in the insertion section were observed.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection, and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation regarding the allegation, the cause was a broken video cable.For the cracked cover glass in the insertion section, the cause was traced to a component failure without any design or manufacturing issues.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|