• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP PRIVATE LABELED FOR CUSTOMER; DBD-ROLLATOR, KNOCK DOWN, WGNS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP PRIVATE LABELED FOR CUSTOMER; DBD-ROLLATOR, KNOCK DOWN, WGNS Back to Search Results
Catalog Number WRX490530
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "when to unlock it from locked and it flopped" and "rollator hand brake lever is just floppy not staying firm in its place.".
 
Additional Manufacturer Narrative
The customer reported that, during use, "when to unlock it from locked and it flopped" and the rollator hand brake lever was "just floppy not staying firm in its place"; no injury occurred.A sample and/or photo was not available, requested information was not received from the customer, finished good/component trending was performed with no trends identified for the specified material number and corresponding defect type, and quality control inspection records were reviewed.The investigation results were not confirmed.A root cause was not determined because a sample was required to determine cause; the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIVATE LABELED FOR CUSTOMER
Common Device Name
DBD-ROLLATOR, KNOCK DOWN, WGNS
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
raveena sampathi
three lakes drive
northfield, IL 60093--275
MDR Report Key50944869
Report Number1417592-2026-54011
Device Sequence Number18754843
Product Code ITJ
UDI-Device Identifier00311917203065
UDI-Public00311917203065
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberWRX490530
Device Lot Number88521010001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-