| Model Number |
CM2000 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from the healthcare provider on 15sep2026 that a right heart catheterization was performed on (b)(6) 2026 to confirm the validity of the pressure sensor readings.The sensor was recalibrated, and the mean was decreased by 9.2 mmhg.Readings were observed under the manufacturer's patient care network database.
|
| |
|
Additional Manufacturer Narrative
|
|
No device analysis results are available because the device remains implanted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Per the instructions for use, right heart catheterization is required for system baseline (mean pressure) calibration and may be needed to recalibrate the baseline (mean pressure) in order to continue to use the system.
|
| |
|