The reported failure was that the product is not working and will not calibrate.Medline received one harmony blood glucose meter in used condition; the sample came with 2 working batteries, test strips and control solutions used by the customer were not sent for investigation, and the sample was tested according to the instructions for use with two tests, one for each control solution.Control solution results were 47 mg/dl for level 1 and 222 mg/dl for level 3, respectively, and no issues were seen during testing; the sample turned on/off as intended, the test strip ejector worked properly, and readings were within control solution range.The complaint was not confirmed, the cause was not confirmed with received sample, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.This is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|