It was reported that when using the bd pyxis¿ medstation¿ 4000 integrated main, drawer 3.2 was failed to open.Customer stated that latch was heard clicking; however, the drawer was not released and unable to pull medication.The customer stated that this malfunction occurred while dispensing the medication which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
|
|
E.1.Initial reporter facility name: (b)(6) hospital.A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 26-jul-2018 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the drawer 3.2 was failed to open.A field service engineer (fse) replaced a damaged u-link plunger, which resolved the reported issue and restored normal operation.As part of the repair, inspected and reseated all cable connections, checked the slide bumpers for proper condition, and cleaned the electronics drawer.Verified functionality after the repair and confirmed the equipment was operating normally.The system functioned as intended after the field service engineer repaired the device.
|