E.1.Initial reporter facility: ascension via christi regional medical st francis campus.A review of the complaint history for (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for (b)(6) was performed from the date of manufacture, 01-feb-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system drawer 1.1 failed.A field service engineer (fse) found that drawer 1.1 had failed and was not displaying any cubies.Hardware test application (hta) testing showed that no drawers were detected.Inspection of the drawer controller board revealed incorrect flashing light patterns.The drawer controller board was replaced, and all row board connections and pyxibus cables were reseated.After the repair, the drawer displayed cubies correctly and was fully functional.Hardware test application testing confirmed the drawer had recovered, resolving the issue.The system functioned as intended after the field service engineer repaired the device.
|