It was reported that when using the bd pyxis¿ medbank tower drawer 01 and drawer 02 cubie drawers failed to open, and the user was unable to issue medications from the affected drawers.The customer stated that there was no delay, but the malfunction occurred when the user tried to issue medication to the patient.However, there were no adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 08-sep-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device's module controller was faulty.A field service engineer (fse) remotely accessed the cabinet and identified cubie drawers in drawer 01 and drawer 02 indicating a yellow status.The user was guided to inspect the drawer cable connections, and the application was restarted; however, both drawers continued to display the yellow indication.Fse performed onsite troubleshooting, verified proper operation of the drawer controllers using a meter, and determined the controllers were functioning correctly.The module controller was replaced, after which remote configuration of the drawers was completed successfully.Final testing confirmed normal drawer operation and functionality.The system functioned as intended after the field service engineer repaired the device.
|