It was reported that when using the bd pyxis¿ medstation¿ es auxiliary tower system hardware failed.The customer tried to restart and recover, but the issue persisted.The customer stated that this malfunction occurred when the user tried to issue medication to patients and caused a delay in dispensing medication to patients.However, there were no adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 23-oct-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system required recovery.A field service engineer (fse) found a failed door icon with no recovery options available.Door 4 was identified as the source of the issue.All doors were unlocked, and the station was rebooted.After the reboot, all doors, including door 4, were successfully recovered.The engineer then accessed the hardware test application (hta) and tested all doors, which passed successfully.The customer performed multiple inventory cycles and confirmed that door 4 was functioning properly, resolving the issue.The system functioned as intended after the field service engineer repaired the device.
|