| Model Number |
500004500000 |
| Medical Device Problem Code |
Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank tower, the cabinet was offline.The customer reported that the pyxis cabinet screen was completely black, preventing access to the system.The customer reported that a malfunction took place when the user tried to dispense medication to the patient.There were no delay or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn(b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn(b)(6) was performed from the date of manufacture, 27-sep-2019 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of this incident, a technical support specialist (tss) advised the user to press the power button located beneath the monitor and found that the display had been turned off.After the display was powered on, guided the caller through logging into the station and verifying functionality.The caller successfully logged in, confirmed the cabinet was operating normally, and verified that the issue was resolved. the system functioned as intended after the technical support specialist guided the user to resolve the issue.
|
| |
|