BOSTON SCIENTIFIC CORPORATION AGENT?; Drug-eluting percutaneous transluminal coronary angioplasty catheter Back to Search Results
Model Number H7493960830300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 08/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a grade c dissection occurred.The 2.75 mm x 20 mm target lesion located at the 2nd diagonal branch was pretreated with two devices with the largest balloon size of 3 mm x 15 mm.The non-complaint balloon had a maximum inflation pressure of 8 and two balloon inflations.Final pretreatment stenosis was 40% with timi flow 3.Ivus with opticross 6 catheter was used to confirmed pretreatment results.The target lesion was then successfully treated with a 3.0 mm x 30 mm agent dcb device.The device achieved an inflation pressure of 6 with an inflation time of 80.However, a grade c dissection was discovered, and an interventional procedure was completed to place a non-boston scientific drug eluting stent.The patient was discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGENT?
Common Device Name
Drug-eluting percutaneous transluminal coronary angioplasty catheter
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
FREUDENBERG MEDICAL
adenauerstrasse 15
wuerselen 52146
GM   52146
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50944990
Report Number2124215-2026-51925
Device Sequence Number18753593
Product Code OOB
UDI-Device Identifier00191506014896
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P230035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberH7493960830300
Device Catalogue NumberH7493960830300
Device Lot Number13464H25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured11/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
Patient Weight138 LB
-
-