A complaint was received, and the customer reported gloves that had a finger missing on the glove.The reported issue was evaluated via the customer-supplied image and returned sample; the vendor was notified and reported that no issues or potential defects could be identified within their batch records and retention samples, and trending did not indicate a trend of "missing finger/pieces" or similar occurring in the field.The reported issue was confirmed, as the missing pointer finger/digit of the glove could be visually observed to be missing, and the product failed to meet specifications.Cause was documented as mfg: machine/automation process error, and the review indicated no adverse trending and no need for further action at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|