| Model Number |
H74939354090 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter issue occurred.The 70% stenosed target lesion was located in the mildly tortuous and moderately calcified left anterior descending (lad) artery.The left main artery was not calcified.An opticross 6 hd imaging catheter was advanced into the left main artery.Resistance was encountered while advancing toward the lad, possibly due to the left main angulation and the catheter began to separate from the wire.When the device was removed, it was noted that the catheter had begun to separate.The procedure was completed using another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
|
| |
|