The reported issue was leaking due to stopcock.A sample consisting of one stopcock attached to the component insert sheet was received; the stopcock was attached to a 20ml syringe filled with tap water to test for leaks, failed due to a visible leak when the plunger was depressed, and a crack in the white valve that directs the fluid path was found.The reported problem was confirmed, and based on the complaint details, the root cause is considered to be a vendor product issue relating to the component manufacturing process.Finished good production records and pack build history were reviewed with no issues identified that would contribute to the reported issue; this complaint has been provided to the component manufacturing facility for further evaluation and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|