Customer reported tubing causing bubbles and high pressure alarm and difficulty manually priming before inserted into pump.Medline quality received 1 ea of dyndtn0007 for evaluation, and visual inspection was conducted with the provided samples.Due to the usage of the device, medline was unable to confirm the complaint; dyndtn0007 is not a pressure-rated set and is to be used only with gravity iv infusion, and the cause code was dc:traced to user - preference/item to spec.A retrospective review of the past two years for this product indicated that there have been no prior complaints in regard to this issue, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|