| Model Number |
SY60WF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A healthcare professional reported during intraocular lens (iol) implant procedure, the surgeon noticed optical part of iol had cracks after the iol was implanted into the eye.The lens was explanted during initial procedure.The procedure was completed on same day.There was no patient harm.Additional information received via regulatory agency stating that the iol appeared normal, with no abnormality observed by the naked eye.After implantation into the patient's eye, a distinct scratch was observed in the central area of the iol under the microscope and procedure was completed by using another unspecified lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|