According to the facility the product was being used in a left heart case; they were performing and lv gram and power injected through the manifold at 600 psi, and the rotating adaptor on the manifold came off the manifold.No sample or photo was received for the investigation; the packs production record could not be reviewed as no lot number was reported, and the bill of materials (bom) and packaging instructions were reviewed with no change to the bom or placement of components that would have contributed to the reported defects.At this time, medline was unable to confirm the reported defects, as no physical samples or detailed photographic evidence was provided to support the investigation; the cause was unconfirmed, with sample required to determine cause.The division will monitor this issue for trending purposes, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|