This vmsr report summarizes 5 malfunction events that occurred during q3 2023 (july 1, 2023, through september 30, 2023).The account reported that five hcs70004rd nebulizers experienced low flow output and failed to meet performance expectations.The reported issue involved devices associated with lot/serial numbers (b)(6).Replacement units were requested.No patient age, weight, sex, race, or ethnicity information was available.No patient or user injuries were reported.The devices were returned for evaluation.
|
|
This vmsr submission is part of a retrospective data remediation review.The included events represent 5 malfunction events that medline became aware of during q3 2023 (july 1, 2023, through september 30, 2023).Associated lot and serial numbers are (b)(6).Udi is (b)(4).Five hcs70004rd nebulizers were evaluated following reports of low flow output and failure to perform as expected.Replacement units were requested for all affected devices.Of the events summarized, all five devices were returned and evaluated.Visual inspection identified filters that were due for replacement on all returned units.Functional testing confirmed the reported malfunction, with flowmeter readings of 2.3 lpm, 0.2 lpm, 0.0 lpm, 0.0 lpm, and 1.2 lpm, respectively, which did not meet product specifications.Internal evaluation identified debris and fluid damage in one unit, debris in multiple units, and a loose piston in one unit.Based on the evaluation, the reported malfunction was confirmed.The cause code was unknown responsible party - cannot determine cause with received sample.Although the malfunction was confirmed through evaluation, a definitive responsible party or root cause could not be established based on the information available from the returned samples.The devices were not labeled for single use, and no devices were identified as reprocessed or reused.No patient age, weight, sex, race, or ethnicity information was available.There were no reports of death, serious injury, medical intervention, follow-up care, or user injury.It has been determined that the reported malfunction could cause or contribute to a serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available, this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|