| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported being unable to complete incoming functional testing.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (unable to baseline) has been confirmed.Upon investigation the monitor was not properly producing a driven ground signal.The cause for the failure was isolated to an open driven ground resistor on the computer/analog board.The root cause for the open resistor could not be positively identified.No adverse event resulted from the damaged monitor.
|
| |
|