It was reported that prior to surgery, the pressure sensor on the fms vue device went off.It was further reported that the high-pressure light came on.This event did not occur during surgery. there were no reports of injuries, medical intervention or prolonged hospitalization. all available information has been disclosed. .
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the product was returned to depuy synthes service center for evaluation.A service evaluation was performed accordance with depuy synthes procedure and the following defects were identified:
functional : intermittent operation, error code: fc004, visual : component damaged.The pump were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The reported complaint was confirmed based on the service evaluation results.The investigation traced the potential cause to faulty parts from wear.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.
|