Catalog Number SSN100255
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "multiple times when administering a vaccine and the needle would come unscrewed from the syringe/immunization and stay in the patients arm or leg, and the syringe in their hand.Also, another time the immunization liquid came out when pushing the plunger through the syringe.".
 
Additional Manufacturer Narrative
The customer reported that when administering a vaccine, the needle would come unscrewed from the syringe/immunization and stay in the patients arm or leg, and another time the immunization liquid came out when pushing the plunger through the syringe; this occurred four times total.Medline received a complaint for item ssn100255, the customer advised that they were using prefilled syringes with the needles, and material trending via sap showed similar complaints during the last 12 months.The investigation results were confirmed, and the documented cause was "comp./part defective, broke." the division has been informed and will continue to monitor trends; no medical intervention, serious injury, or follow-up care has been reported.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
NEEDLE,HYPODERM,SAFETY, 25GX1
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rizwan shaik
three lakes drive
northfield, IL 60093--275
MDR Report Key50945379
Report Number1417592-2026-54066
Device Sequence Number15827390
Product Code FMF
UDI-Device Identifier30193489031984
UDI-Public30193489031984
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K213054
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSSN100255
Device Lot Number89721070001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/24/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2024
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-