Synthes GmbH 2.7MM CANNULATED DRILL BIT/QC 160MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
310.670 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that in the surgical act, when the 1.25 mm guide needle broke while the doctor was using it.A fragment of this needle was stuck together with bone tissue inside the 2.7 mm drill bit, blocking its canal.The surgeon, the scrub nurse and i tried to remove it on the operating table but it was not possible to uncover it, so the drill bit was left clean externally and marked with red tape.This generated a small delay while the fragment was removed, without any discomfort on the part of the doctor, who even collaborated in the process.The eventuality was resolved by requesting another replacement 1.25 mm needle from the institution, successfully placing the last 3.5 mm screw and completing the surgery satisfactorily.There was no patient harm or surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4)
this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|