The account reported the needle was out of packaging, resulting in a needle stick, and no medical intervention, serious injury, or follow up care was reported.No sample or photo was received for the investigation; finished good production records were reviewed with no non-conformances noted, the bill of materials was reviewed, and one each of component #104163 is part of subassembly (s01-02) and is placed inside a poly bag along with several other components.The reported defect could not be confirmed as there were no samples or detailed pictures received, and without a sample it is difficult to determine the true root cause.The supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|