|
The reported product problem/issue was leakage, with the account reporting leak observed from bonding joint between the tubing and the female connection.No stock check was required, as this was a sample build, and no sample was received for evaluation; a 6-month review of complaint history revealed 0 similar complaints reported by the customer account and 0 similar complaints reported against the finished good part number.Based on the investigation results, the reported event was not able to be confirmed at this time as there was no sample returned for evaluation.Root cause remains inconclusive at this time as the reported defect was not able to be confirmed, and medline will continue to monitor for additional occurrences and adverse trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|