Per the end user's report, the backup battery failed during the recent power outage.Medline received one battery only in used condition; upon further investigation of the sample, when tested there was no power from the battery, with no damage to the cord or plug end, and pic testing resulted in a functional defect.The investigation results were confirmed; it is possible that this may be due to an inner electrical issue, and the cause was documented as cannot determine cause with received sample.At this time the quality complaint has been confirmed, and the division will monitor through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.This is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|