The account reported finding leaking stopcock.A sample was received for the investigation; three samples were evaluated, finished good production records were reviewed, and the finished good pack build was reviewed.During the testing of the samples no leakage or damage was observed, and the reported issue was unable to be replicated at this time.Though medline could not confirm with the physical samples, the reported issue is similar in nature to a recent increase in complaints for the component, and the nature of the damage indicates that the error occurred due to a potential molding defect; the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|