The customer expressed that the lift is not functioning properly; troubleshooting notes state that the lcd/control box do not appear to be receiving power, the control box output was measured at 0v, the led does not come on when the charger is plugged in, and the lift is still not working.Medline reviewed the customer-provided photo and noted the product appears to be in used condition with dirt buildup visible on the lift base; a physical sample was not provided, and a returned scale was inspected and found to be an incorrect replacement part not compatible with the sit-to-stand lift.The investigation results were not confirmed: the reported issue with the control box cannot be identified from the photo, and the control box sample has not yet been received.No root cause was identified, and the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.This is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|