| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported being unable to complete incoming functional testing.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (cannot run detect and treat testing due to functional issue) has been confirmed.Upon investigation the monitor was unable to complete incoming functional testing.The cause for the failure was isolated to the defibrillator pca and computer/analog board.High voltage travelled to low voltage circuitry, damaging multiple components on the pcas.Additionally, the rear response button connector was unplugged causing faults.The root cause for the high voltage travelling to low voltage circuitry and disconnected cable could not be positively identified.No adverse event resulted from the damaged monitor.
|
| |
|