| Model Number |
DS2110X11B |
| Medical Device Problem Codes |
Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy evo ventilator was returned to the manufacturer's service center for evaluation.There was no harm or injury reported.During the evaluation, a ventilator inoperative alarm was observed indicating data in eeprom was corrupt on the system/cpu.The system printed circuit assembly (pca) needed to be replaced to correct the issue.A final report is being submitted.If new information becomes available that changes the outcome of the investigation, a follow up/supplemental report will be completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of ventilator inoperative alarm indicating data in eeprom was corrupt on the system/cpu is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device logs were not available; therefore, the date of the most recent ventilator inoperative occurrence could not be determined.Accordingly, the date the device was evaluated has been used as the event date for reporting purposes.Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Search Alerts/Recalls
|