| Model Number |
BF-H190 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Material Fragmentation (1261); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that the scope experienced a scope communication error e216.The issue was found during set up for an unspecified procedure.No harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: chip on the edge of the distal end cover and static screen noise due to error
a root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunctions: expected or random component failure without any design or manufacturing issue
should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|