The customer reported that the pressure bag is not staying inflated and over the course of 30 minutes required reinflation for invasive monitoring in a critically ill patient.No sample or photo received for our investigation, and the production records could not be reviewed since a lot number was not provided.Based on the complaint details and our investigation, the complaint is not confirmed.The investigation was limited without receipt of the lot number, physical sample or detailed photographic evidence from the customer; the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|