| Model Number |
161001 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hamilton medical ag received the following event description (summary):
it was reported that, during an inspection on 21 august 2026, the self-test of a hamilton-c1 ventilator, s/n (b)(6) failed and technical faults (tf) 431002 and 444004 were displayed.No patient, user, or third party was involved.No harm or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Hamilton medical ags reference: (b)(4);
hamilton medical ags conclusion:
it was reported that a hamilton-c1 (pn 161001, s/n (b)(6)) failed the self-test and displayed technical fault (tf) 431002 (blowerdisconnected) and tf 444004 (voltageoutoftolerance).No patient involvement, harm, or medical intervention was reported.
investigation showed repeated occurrences of tf 431002 and tf 444004.The reported condition was attributed to a control-board-related electronic malfunction and was no longer observed after replacement of the control board.
hamilton medical ag considers this case closed.
|
| |
|