| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported being unable to complete incoming functional testing.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (cannot run detect and treat testing due to functional issue) was confirmed.Upon investigation the monitor was unable to power on.The cause for the failure was isolated to multiple defective components on the computer/analog pca.The root cause for the defective components could not be positively identified.No adverse event resulted from the damaged monitor.
|
| |
|