• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. ETHICON SECURESTRAP; Staple, Implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. ETHICON SECURESTRAP; Staple, Implantable Back to Search Results
Catalog Number STRAP25R
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and absorbable straps were used.The product clip unable to stay in position.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 9/30/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3: investigational summary: the product was returned to eth for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the strap25r device was returned with no apparent damage.In addition, the foil packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device was manually activated; however, during subsequent actuations, no straps were deployed despite multiple firing attempts.The instrument was subsequently disassembled for further evaluation.Upon internal inspection, the fork prongs were found to be deformed, preventing proper advancement and deployment of the straps.Additionally, twelve (12) straps were observed remaining within the cannula shaft.The deformed fork prongs are considered the most likely cause of the reported firing failure.The event reported was confirmed and it is related to improper use of the device.A possible cause for the condition of the prongs of the fork deformed is indicative of the device being fired into bone or another hard surface, thus rendering device unusable.A minimum tissue thickness of 6.7 mm is required when applying the fastener over underlying bone, vessels, or viscera.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHICON SECURESTRAP
Common Device Name
Staple, Implantable
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-Juarez
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50945845
Report Number2210968-2026-09917
Device Sequence Number13149649
Product Code GDW
UDI-Device Identifier10705031470552
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K093845
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSTRAP25R
Device Lot Number10AZ7R
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/22/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-