| Model Number |
CCAET0 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
09/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, the leading haptic was folded backwards and stuck to back of optic; pulling it loose with a sinskey instrument resulted in a peripheral posterior capsule defect.Additional information has been requested but is not available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|