This vmsr submission is part of a retrospective data remediation review.The included events represent 3 malfunction events that medline became aware of during july 1, 2025, through september 30, 2025.Associated lot numbers 29524110006, 29524060008, and 29525040007.The quantity associated with the reported issue was 6 bx.A complaint was received regarding mds202105 following reports that wooden applicator sticks were splintering within the package prior to use.Physical samples associated with lot numbers 29524110006, 29524060008, and 29525040007 were returned for evaluation.Medline quality received 6 bx in total, consisting of 4 bx from lot 29525040007 and 1 bx each from lots 29524110006 and 29524060008.Visual inspection identified no abnormalities with the packaging or product.All boxes were sealed and unopened at the time of receipt.Fifteen sticks were examined from each box, totaling 90 ea evaluated.No splintered sticks were identified during inspection.Additionally, the sticks did not break easily when removed from the packaging and required a reasonable amount of force to fracture.Based on the investigation, the reported issue could not be confirmed, and no evidence was identified indicating that the product failed to meet specifications.Complaint trending conducted over the previous two years identified only one similar complaint, which was also unconfirmed and associated with an unknown lot.No capa, scar, or additional risk management activities were deemed necessary based on the available information.The complaint will continue to be monitored through routine trend analysis.6 bx were returned for evaluation.The devices were labeled for single use.No devices were identified as reprocessed or reused.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|