• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX Pro; system, appendage closure, left atrial

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX Pro; system, appendage closure, left atrial Back to Search Results
Model Number M635WU60350
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Stroke/CVA (1770); Muscle Weakness/Atrophy (1967); Visual Disturbances (2140); Loss of consciousness (2418); Thromboembolism (2654); Speech Disorder (4415)
Date of Event 08/31/2026
Type of Reportable Event Death
Event or Problem Description
It was reported a stroke and patient death occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted.Immediately following the procedure, the patient exhibited right-sided weakness, inability to awaken, absence of speech attempts, inability to follow commands, and dysconjugate gaze.The patient was subsequently diagnosed with a stroke and was admitted to the hospital.In response to the stroke, thrombectomy was performed in interventional radiology to remove the clot and mechanical ventilation were performed.Four days following the index procedure, the patient died.The reported cause of death was thrombosis of the left middle cerebral artery (mca) with a large left mca infarct and significant cerebral edema resulting in a 1.8 cm left-to-right midline shift.The physician believed that a combination of preexisting comorbidities and the concomitant procedure led to the stroke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX Pro
Common Device Name
system, appendage closure, left atrial
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50945914
Report Number2124215-2026-51941
Device Sequence Number21216027
Product Code NGV
UDI-Device Identifier00191506004620
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberM635WU60350
Device Catalogue NumberM635WU60350
Device Lot Number0039047488
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured04/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Death; Hospitalization; Other;
Patient Age80 YR
Patient SexFemale
-
-