Catalog Number MPH6540
Medical Device Problem Codes High Readings (2459); Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "their checks in the morning are consistently a little out of range.The level 1 is always just slightly higher than range, and the level 3 is just slightly lower than normal range.".
 
Additional Manufacturer Narrative
According to the customer contact, when using the blood glucose monitoring system, they noticed that their checks are consistently out of range; the level 1 is slightly higher and the level 3 is slightly lower than normal.Medline received 1 ea harmony blood glucose monitor; visual inspection found no cosmetic damage, and functional testing with new batteries and a controlled solution test produced l1-34 mg/dl and l3-205mg/dl, both within acceptable ranges.The complaint was not confirmed.Per the customer contact, this was due to user error, with no patient harm to report and no patients or rns affected; no trends were identified for the specified material number and its corresponding defect type, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER, GLUCOSE, HARMONY,PROF.USE ONLY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joseph weiske
three lakes drive
northfield, IL 60093--275
MDR Report Key50945961
Report Number1417592-2026-54163
Device Sequence Number10451587
Product Code LFR
UDI-Device Identifier10888277675674
UDI-Public10888277675674
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K161299
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH6540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/27/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-