• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Synthes GmbH 5MM HEX FLEXIBLE SCREWDRIVER; Orthopedic Manual Surgical Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH 5MM HEX FLEXIBLE SCREWDRIVER; Orthopedic Manual Surgical Instrument Back to Search Results
Catalog Number 03.037.028
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the driving cap f/insertion handle, multihole drill sleeve, trocar yell, scrdriver-hex 5 flex cann, extract-instr f/tfna helical blade+scr, scrdriver short xl25, scrdriver w/quick-coupl hex 12 short xl2, hand lrg cann w/quick-coupl hex12, connecscr f/tfna helical blade+scr, and scrdriver shaft self-holding l165 f/star were associated with wear-related and deformation issues.The reported failures included stripped or worn threads that did not engage securely, deformed holes that made guide wire passage difficult, a bent trocar that did not fit the cannula properly, dull screwdriver features that did not seat securely or function properly, and a handpiece that no longer clicked into screwdriver shafts.There was no patient involvement.The complaint devices were retained by the customer.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM HEX FLEXIBLE SCREWDRIVER
Common Device Name
Orthopedic Manual Surgical Instrument
Manufacturer (Section D)
Synthes GmbH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50945962
Report Number8030965-2026-09944
Device Sequence Number18754858
Product Code LXH
UDI-Device Identifier10886982070449
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.037.028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-